Technical Specifications
Parameter |
Details |
System Type |
Random access, fully automatic, discrete, STAT priority |
Throughput |
Constant speed 800T/H for colorimetric test, 400T/H for ISE |
Light Source |
Long life halogen lamp |
Wavelength |
340–800nm, 12 wavelengths |
Analysis Method |
End - point, kinetics, fixed - time, etc. |
Calibration Method |
1 point method, 2 point method, multiple point linear method, non - linear method |
Probe |
Independent sample probes, reagent probes, automatic liquid level detection, probe liquid level tracing and clot detection function |
Rinsing Mechanism |
Probe inner wall high pressure rinsing |
Mixing Mechanism |
2 independent mixers ensure sufficient reaction |
Reaction Cuvette |
160 new type reaction cuvettes |
Temperature Control |
The temperature of reaction disk incubation bath is 37°C±0.1°C |
Sample Tube |
140 sample positions, supporting multiple tubes and sample cups |
Reagent Position |
2 compartments, 132 reagent positions and 2 detergent positions in total, with continuous refrigeration; supporting at most 4 kinds of reagents test function |
Sample Volume |
1.5 μl–35μl, 0.1μl increment |
Reagent Volume |
15 μl–350μl, 1μl increment |
Reaction Volume |
Minimum reaction volume 120 μl |