PRODUCT

PRODUCT

Fully Automated Nucleic Acid Extraction Instrument - Auto-Pure 4800

The 4800M/2400M contains a temperature module that prevents cross-infection, and with the corresponding software, the whole process of methylation conversion can be automated, making complicated experiments simpler. In addition, the full-coverage UV disinfection system and negative pressure filtration system effectively prevent cross-contamination. As a high-efficiency intelligent workstation, Auto-Pure series fully automated nucleic acid extraction and purification instrument can bring you a superior product experience.

Product Detail

Test Menu

Product Tags

Auto-Pure 4800 全自动核酸提取纯化仪.png
Auto-Pure series automatic nucleic acid extraction and purification instrument is a high-throughput automatic nucleic acid extraction equipment, the instrument is based on the magnetic bead method to extract nucleic acids, the purification process of pipetting, lysis, washing, elution, the construction of the PCR system and other steps are automatically completed in the workstation, one-stop operation, without human intervention, and can be extracted at the same time 48X1ml (4800/4800M) or 24X10ml (2400/2400M) sample system, compared with the manual extraction, efficient and convenient at the same time, standardised, automated operation steps to avoid human error. It can extract 48X1ml (4800/4800M) or 24X10ml (2400/2400M) samples at the same time. Compared with manual extraction, it is highly efficient and convenient, while the standardised and automated operation steps can avoid human errors, which makes the results of the experiments more reliable and more reproducible.

Technical Parameters

Model No.: Auto-Pure 4800/Auto-Pure 4800M Automatic Nucleic Acid Extractor

Functions: automatic sample filling, nucleic acid extraction and PCR reaction system construction.

Information management: support for external USB scanning gun, the sample area comes with barcode scanning function, inserted into the sample rack automatic scanning entry

Pipetting principle: air displacement pipetting technology

Number of pipettes: two for single lane 1000u, one for single lane 200μL.

Pipetting range: 1-1000μL

Matching tips: our tips, and other compatible with the instrument's domestic / imported tips

Liquid detection: Pipette gun with integrated pneumatic liquid level detection, tip clogging detection and residual liquid volume detection.

Number of samples extracted at one time: 1-48

Nucleic acid extraction volume: 50-1000μL

Extraction uniformity: CV ≤ 5%

Single sample consumables: support

Extraction temperature control: elution, lysis temperature control: room temperature ~ 120 ℃.

Reagent refrigeration module: optional, 0℃~105℃

Temperature control module (Auto-Pure 4800M): 4℃~105

UV sterilisation: extraction area, pipetting area and disposal area are independently equipped with UV sterilisation function.

Fan Filtration: Equipped with three HEPA high-efficiency filters

Instrument interface: USB interface

Power supply: 220V,8A,50/60 Hz

Power: standby 30W, maximum 550W

Noise: ≤60dB

Net weight: 130kg