Technical Parameters
Model: Leap-Pure R16
Function: Fully automatic pipetting and nucleic acid extraction
Extraction principle: based on the magnetic bead method, the combination of the magnetic rod and pipetting extraction method
Sample throughput: 1-16
Applicable samples: FFPE samples, whole blood, plasma, faeces and sputum, etc.
Reagent pre-packaging: Support pre-packaged reagents directly on the machine, automatic membrane piercing.
Information management: Support external code scanner for sample code scanning.
Visual identification: optional visual module, automatic identification of the number of consumables and specifications
Sample processing system: 1-15mL
Minimum elution volume: 20μL
Difference between purification wells: CV ≤ 5%
Software operation: 10-inch colour large touch screen, windows system software.
Account management: authority hierarchical management, procedures and instruments can be customised according to the authority settings
Instrument interface: USB port, LAN port
Contamination control: built-in HPEA components, UV disinfection
Power supply: AC220V, 50Hz, rated power 450W
Extraction Temperature Control [Usage]: For incubation in the extraction process, such as lysis, elution, etc.
Extraction Temperature Control [Temperature Range]: 40℃~120
Extraction Temperature Control [Temperature Accuracy]: ≤±1℃ @ 100℃
Extraction Temperature Control [Temperature Uniformity]: ≤3℃ @ 100℃
Extraction temperature control [Temperature rise speed]: ≤5 min
Reagent refrigeration [Use]: for reagent refrigeration, support power-on automatic refrigeration
Reagent Refrigeration [Temperature Control Range]: 4℃~95℃.
Reagent Refrigeration [Temperature Accuracy]: ≤±1℃ @4℃.
Reagent Refrigeration [Temperature Uniformity]: ≤1℃ @4℃
Pipetting module 1 [Application]: Nucleic acid extraction
Pipetting Module 1 [Pipetting Accuracy]: 20μL:±8% 100μL:±2% 500μL:±1% 1000μL:±1
Pipetting Module 1 [Pipetting Accuracy]: 20μL ≤ 2.5% 100μL ≤ 1.5% 500μL ≤ 1% 1000μL ≤ 1%
Pipetting Module 1 [Pipetting Range]: 20-1000μL
Pipetting module 2 [Application]: PCR system construction
Pipetting Module 2 [Pipetting range]: 1-200μL
Pipetting Module 2 [Pipetting Accuracy]: 1uL≤±12% 20μL≤±2% 100μL≤±1% 100μL≤±1
Pipetting Module 2 [Pipetting Accuracy]: 1μL ≤5% 20μL ≤2% 100μL ≤1% 200μL ≤1%