Technical Parameters
Model: Leap-Pure S24/H24
Principle of operation: magnetic bead method, magnetic rod, pipetting/lance blowing mixing
Applicable samples: FFPE samples, whole blood, plasma, faeces, sputum, etc.
Consumables: 2mL reagent strip/8mL reagent strip/15mL reagent strip
Sample throughput: 1-24
Sample processing system: S24: 1mL-15mL H24: 1mL-20mL
Elution volume: 20-200μL
Pipetting range: 20-1000μL
Pipetting accuracy: 20μL: ≤ 5% 100μL: ≤ 1.5% 500μL: ≤ 1% 1000μL: ≤ 1%
Pipetting Accuracy: 20μL: ±8% 100μL: ±2% 500μL: ±1% 1000μL: ±1
Difference between purification wells: CV ≤5%
Temperature control range: 37℃~120
Temperature increase speed: ≤5 min (40℃~100℃, 25℃ environment)
Temperature accuracy: lysis binding wells: ±1℃, elution wells: ±1℃, @100℃ (25℃ environment)
Temperature uniformity: cleavage binding wells: 3℃, elution wells: 3℃, @100℃ (25℃ environment)
Software operation: 10-inch colour large touch screen, Windows operating system
Account management: permission hierarchical management, procedures and instruments can be customised according to the permissions settings
Instrument interface: USB port, LAN port
Vision system: optional vision system, the position of the consumables specifications automatically confirmed
Information management: support for external scanner for sample code scanning
Contamination control: built-in HPEA components, UV disinfection
Power supply: AC220V, 50Hz, rated power 450W.
Overall dimensions (W×D×H): 740mm×640mm×640mm
Weight: 85kg