Technical Parameter
|
Model |
Auto-Pure 2400 Automatic Nucleic Acid Extraction and Purification Instrument |
|
Functions |
automatic sample/reagent loading, nucleic acid extraction and PCR reaction system construction. |
|
Information management |
support for external USB scanning gun, the sample area comes with barcode scanning function, inserted into the sample rack automatically scanning entry. |
|
Pipetting principle |
air displacement pipetting technology |
|
Number of pipettes |
5000μL single lane, 200μL single lane. |
|
Pipetting range |
1-5000μL |
|
Matching Tips |
Our tips and other domestic/imported tips compatible with this instrument. |
|
Liquid detection |
Pipette gun with integrated pneumatic liquid level detection, tip clogging detection and residual liquid volume detection. |
|
Number of samples extracted at one time |
1-24 |
|
Nucleic acid extraction volume |
50-10000μL |
|
Extraction uniformity |
CV ≤ 5% |
|
Single sample consumables |
support |
|
Extraction temperature control |
elution, lysis temperature control: room temperature ~ 120 ℃. |
|
UV sterilisation |
UV sterilisation function is independently set up in extraction area, pipetting area and waste area |
|
Fan Filtration |
Equipped with three HEPA high efficiency filters |
|
Instrument interface |
USB interface |
|
Power supply |
220V,8A,50/60Hz |
|
Power |
standby 30W, maximum 600W |
|
Noise |
≤60dB |
|
Net weight |
130kg |